FAULT
|
ERROR CODE
|
WHAT IT MEANS
|
HVPS
HVPS
|
201
202
|
BOOST FAULT, POWER SUPPLY
B+ AND B+/2 ARE MORE THAN 20 V APART
|
HVPS
|
211
|
THE BLEED_FAULT_Q IS SET
|
HVPS
|
213
|
IGBT_FAULT_Q IS SET
|
SAFETY
|
214
|
CROWBAR_FB_Q IS SET
|
HVPS
|
215
|
DC_FAULT_Q IS SET
|
HVPS
TEMPERATURE
|
217
221
|
UNEXPECTED POWER
SETP2 IS ACTIVE, WATER TEMP >65 DEGREES C
|
TEMPERATURE
|
224
|
TIPTEMP IS GREATER THAN 20 (LASER) OR 30 (FLASH) DEGREES C IN READY OR EXPOSURE
|
SPOTSIZE
|
225
|
SPOTSIZE CHANGED DURING EXPOSURE
|
PARK SENSORS
|
226
|
SELECTED HP WAS PARKED DURING EXPOSURE
|
TEMP SENSOR
|
229
|
WIRES TO TEMP SENSOR ARE CONNECTED OPPOSITE.
|
ENERGY FLUENCE
|
230
|
MAIN DETECTOR ACCUMULATED ENERGY IS LESS THAN 50% OF THE REQUESTED ENERGY
|
ENERGY
|
231
|
MAIN DETECTOR ACCUMULATED ENERGY IS MORE THAN 150% OF THE REQUESTED ENERGY
|
ENERGY
|
232
|
ERROR LAMP OVER ENERGY
|
ENERGY
|
235
|
FLASH HP EFFICIENCY IS LOW – REPLACE HP SOON
|
ENERGY
|
237
|
FIREPULSE FAULT – NOT ALL MICROPULSES WERE DELIVERED OR THE FOOTSW WAS RELEASED DURING TITAN EXP
|
REFERENCE VOLTAGE
|
240
|
REFERENCE VOLTAGE IS LESS THAN 80% OR GREATER THAN 120% OF EXPECTED (10V)
|
REFERENCE VOLTAGE
|
241
|
REFERENCE VOLTAGE IS LESS THAN 80% OR GREATER THAN 120% OF EXPECTED (VCC)
|
REFERENCE VOLTAGE
|
242
|
REFERENCE VOLTAGE IS LESS THAN 80% OR GREATER THAN 120% OF EXPECTED (5 V)
|
LAMP
|
243
|
AFTER THE LASER LAMP IS LIT, THE LAMP VOLTAGE GOES OUT OF RANGE
|
REFERENCE VOLTAGE
|
244
|
REFERENCE VOLTAGE IS LESS THAN 80% OR GREATER THAN 120% OF EXPECTED (15 V)
|
LAMP
|
246
|
FLASH HP LAMP FAILED TO START
|
LAMP
|
248
|
THE FLASH LAMP CURRENT SENSOR REPORTS NO CURRENT
|
LAMP
|
249
|
LASER LAMP FAILED TO START
|
TRIGGER SWITCH
|
250
|
TRIGGER SWITCH (HAND BUTTON OR FOOTSWITCH) IS MALFUNCTIONING
|
SHUTTER
|
251
|
SHUTTER WON’T OPEN
|
SHUTTER
|
252
|
SHUTTER WON’T CLOSE
|
COMMUNICATION
|
270
|
NO RESPONSE FROM DISPLAY BOARD IN 400 MS
|
SHOTS EXPIRED
|
288
|
SHOTS EXPIRED, MUST RESET CHIP.
|
FOOTSWITCH USE
|
289
|
THE FOOTSWITCH WAS RELEASED DURING A TITAN PULSE
|
CAPACITY
|
293
|
1000 SHOTS LEFT IN HP
|
CAPACITY
|
294
|
2000 SHOTS LEFT IN HP
|
LAMP
|
301
|
HVPS WON’T BOOST – CHECK FUSE
|
HVPS
|
303
|
TITAN HPP MODE FAULT
|
LAMP
|
310
|
HVPS FAULT
|
HVPS
|
312
|
BOOST_OVP_Q IS SET
|
HVPS
|
316
|
I_LMP_FAULT_Q IS SET
|
HVPS
|
317
|
INADVERT_EXP_Q IS SET
|
HVPS
|
318
|
MODE_FAULT_Q IS SET
|
SHUTTER
|
319
|
SHUTTER_FAULT_Q IS SET
|
WATER FLOW
|
322
|
PRESSURE SWITCH IS NOT CLOSING WHEN PUMP IS ENGAGED.
|
TEMPERATURE
|
323
|
S/W CANNOT READ TEMPERATURE OF THE LASER HANDPIECE.
|
TEMPERATURE
|
327
|
S/W CANNOT READ TEMPERATURE OF THE FLASH HANDPIECE.
|
LOW EFFICIENCY
|
332
|
HEAD POWER IS LOW.
|
DETECTOR MISMATCH
|
333
|
MAIN DETECTOR SIGNAL MISSING – CHECK CONNECTIONS – W/E367 CHECK FOR DETECTOR LEAKAGE
|
DETECTOR MISMATCH
ENERGY
|
334
335
|
LASER MAIN AND SAFETY ENERGY DETECTORS DO NOT AGREE.
LIMELIGHT OR PROWAVE EFFICIENCY IS LOW.
|
ADC FAULT
|
336
|
ANALOG TO DIGITAL CONVERTER FAILED
|
FIREPULSE FAULT
|
337
|
THE EXPOSURE WAS ABORTED BEFORE ALL MICROPULSES WERE DELIVERED
|
SAFETY
|
338
|
DETECTOR INTEGRATOR GAIN HARDWARE FAULT – CHECK SWITCHES IN SW2 BLOCK
|
DETECTOR MISMATCH
|
339
|
FLASH HP MAIN AND SAFETY ENERGY DETECTORS DO NOT AGREE.
|
LAMP
|
345
|
THE FLASH LAMP WENT OUT AFTER A PULSE – LAMP MIGHT BE BROKEN
|
LAMP
|
347
|
THE FLASH LAMP CURRENT SENSOR REPORTS CURRENT WHEN THE LASER LAMP IS ON
|
NVRAM
|
360
|
NVRAM CHECKSUM DOESN’T MATCH OR NVRAM VERSION NUMBER DOESN’T MATCH
|
CALIBRATION
|
361
|
DURING THE TEST FIRE, THE SYSTEM DID NOT MEET MINIMUM EFFICIENCY LIMIT.
|
CALIBRATION
|
362
|
ANY ERROR OCCURRED DURING POST LIGHT REG CONTROLLED TEST FIRE
|
PROM
|
363
|
PROM CHECKSUM DOESN’T MATCH
|
SAFETY
|
364
|
POST FIRED THE CROWBAR AND THE CROWBAR_FB_Q IS NOT SET AFTER 20 MS
|
SAFETY
|
365
|
TIPPIC FAILED ITS POST – MOST LIKELY LEAKAGE CURRENT IN THE TIP THERMISTOR
|
NVRAM
|
366
|
THE SERVICE MODE CALIBRATION WAS NOT COMPLETED
|
SAFETY
|
367
|
POST SAFETY CIRCUIT TEST FAILED
|
NVRAM
|
368
|
THE TOUCH SCREEN CALIBRATION WAS NOT COMPLETED
|
CALIBRATION
|
369
|
POST TEST FIRE DID NOT FINISH IN 20 MS
|
COMMUNICATION
|
371
|
HANDPIECE COMMUNICATION PROTOCOL FAILURE
|
CALIBRATION
|
372
|
MODEL NOT PROGRAMMED TO EITHER XEO OR GENESIS PLUS
|
COMMUNICATION
|
380
|
COMMUNICATION TIMEOUT TALKING TO THE FLASH HP
|
CALIBRATION
|
381
|
DATA IN THE FLASH HP IS UNRECOGNIZED
|
CALIBRATION
|
382
|
FLASH HP TYPE IS UNKNOWN
|
COMMUNICATION
|
383
|
FLASH HP SHOT COUNT INFO HAS A CORRUPTED DIGITAL SIGNATURE – RESAVE SHOT DATA USING SERVICETOOL
|
COMMUNICATION
|
384
|
COMMUNICATION FAILURE TALKING TO THE FLASH HP
|
COMMUNICATION
|
385
|
FLASH HP IBUTTON DATA CRC FAILS
|
CALIBRATION
|
386
|
XEO HP IS NOT AUTHORIZED ON THIS MACHINE – CHECK ENABLED HPS WITH MODELMAKER
|
CALIBRATION
|
387
|
XEO HP IS NOT CALIBRATED
|
SOFTWARE
|
390
|
ASSERTION FAILURE – SOFTWARE LOGIC PROBLEM
|
RELAY
|
391
|
XEO HP RELAY FAULT
|
SOFTWARE
|
392
|
SOFTWARE VERSION MIS-MATCH … CHECK THE DISPLAY FLASH / U5 VERSIONS FOR CORRECTNESS
|